1. Theory
- Definition: Kelvin probe force microscopy measures the contact potential difference between an AFM tip and the sample by applying a DC bias that nulls the electrostatic force or force gradient at a modulation frequency (Bruker; Panchal et al.).
- Work function: With a tip calibrated on a reference, the CPD gives the sample’s work function or surface potential (Panchal et al.; Bruker).
- AM and FM detection: Amplitude modulation senses the force and averages over the cantilever; frequency modulation senses the force gradient and localizes the signal at the apex (Panchal et al.; Bruker).
- Related methods: EFM measures the phase shift from electrostatic force gradients without nulling (Panchal et al.).
2. Working Principle
- Modulation: An AC bias at frequency creates a force at proportional to (Bruker).
- Feedback: A second loop adjusts to null that component; the null value maps the CPD (Bruker; Panchal et al.).
- Lift or single pass: Two-pass lift mode separates topography and potential; single-pass FM and PeakForce modes measure both together (Bruker).
- Calibration: Tips are calibrated on Au, HOPG or Au/Al references, often with UPS values (Panchal et al.; Bruker).
3. Applications
- Semiconductors and devices: Kelvin-voltage contrast across p-n junctions versus doping, and surface band bending (Polak and Wijngaarden).
- 2-D materials: Graphene layer thickness from surface potential (Panchal et al.).
- Solar cells: Photovoltage and work-function changes under illumination (Bruker).
- Metals and alloys: Phase contrast in Sn-Pb and corrosion studies (Bruker).
Electrostatic Detection
| Name | Formula | Related Formulae | Variables/Constants |
|---|---|---|---|
| Electrostatic force | |||
| Contact potential and work function | |||
| Frequency-modulation KPFM |
Resolution and Accuracy
| Name | Formula | Related Formulae | Variables/Constants |
|---|---|---|---|
| Lateral resolution | |||
| Measured-potential accuracy | |||
| Repeatability on a reference |
Applications and Environment
| Name | Formula | Related Formulae | Variables/Constants |
|---|---|---|---|
| Illumination and charge | |||
| Humidity and environment |
Important Questions
Ques: How does KPFM measure the contact potential difference?
By nulling an electrostatic force. Tip and sample form a capacitor; with a bias , the force contains a component at proportional to and one at proportional to (Bruker; harmonics computed). A lock-in detects the component and a feedback loop adjusts until it vanishes, at which point (Bruker). Since the CPD is the work-function difference divided by e, the map shows local work function or surface potential. The largest possible CPD is the 3.79 between osmium and caesium (Bruker; computed), so semiconductor signals of tens to hundreds of millivolts require millivolt precision. Reference: Bruker, ‘PeakForce Kelvin Probe Force Microscopy’, application note AN140 (several equations garbled) (fetch-tool summary); L. Polak and R. J. Wijngaarden, ‘Two competing interpretations of Kelvin probe force microscopy on semiconductors put to test’, arXiv:1605.05539: V_CPD = (W_s - W_p)/e with feedback voltage positive when applied to the sample, surface band bending from surface states or fixed charge, Kelvin-voltage difference across Si p-n junctions versus doping, surface dipoles from termination or adsorbed molecular layers, ions from the environment depositing fixed surface charge, measurements in air (fetch-tool summary).
Ques: Why is FM-KPFM more accurate and sharper than AM-KPFM?
Because it senses the force gradient, which is concentrated at the tip apex. AM-KPFM responds to the force on the whole cantilever and cone, so it averages over a wide area: Panchal et al. give 50 to 70 resolution for AM against below 20 for FM, and Bruker estimates that at 5 lift half the FM signal comes from within 15 of the apex, about 10 resolution. Averaging also reduces contrast: on a known 4 step, AM read 27.6 per cent low while FM read 4.4 per cent high (Panchal et al.), and on Sn-Pb, AM gave 97 against 235 for FM (Bruker), 59 per cent lower (computed). Resolution degrades with lift, to about 200 at 50 (Bruker). Reference: V. Panchal, R. Pearce, R. Yakimova, A. Tzalenchuk and O. Kazakova, ‘Standardization of surface potential measurements of graphene domains’, arXiv:1305.0465 (NPL) (fetch-tool summary); Bruker, ‘PeakForce Kelvin Probe Force Microscopy’, application note AN140 (several equations garbled) (fetch-tool summary).
Ques: What is the potential difference between single- and bilayer graphene?
Panchal et al. measured 50 with AM-KPFM, 150 with FM-KPFM and 110 with electrostatic force spectroscopy, and cite literature values of 135 in vacuum, 25 in air, 120 and 68 . The AM value is low because of spatial averaging: it is 67 per cent below the FM value (computed), consistent with AM’s 27.6 per cent underestimate of a known step. Humidity also matters, since water gating changes the values and the cited vacuum figure is 5.4 times the air figure (computed). The FM value of about 150 , or the vacuum 135 , is the more likely intrinsic difference, while air and AM values understate it (my reading). Reports should therefore state mode and environment. Reference: V. Panchal, R. Pearce, R. Yakimova, A. Tzalenchuk and O. Kazakova, ‘Standardization of surface potential measurements of graphene domains’, arXiv:1305.0465 (NPL) (fetch-tool summary).
Ques: How is the tip calibrated, and are Panchal et al.’s graphene work functions consistent?
Calibration measures the CPD on a reference with known work function. Panchal et al. took gold at 4.82 from UPS and derived a probe work function of 4.09 , then reported 4.55 and 4.44 for single and bilayer graphene from CPDs of 0.20 and 0.35 . These numbers do not all fit one relation: adding the CPDs to 4.09 gives 4.29 and 4.44 , matching bilayer but not single layer, while subtracting gives 3.89 and 3.74 (computed). The CPD difference, 0.15 , matches their FM layer contrast, so the relative values are more likely reliable, and the single-layer absolute value should be treated with caution until checked against the original paper (my reading). Polak and Wijngaarden define ; instruments that bias the tip instead reverse the sign (my reading). Reference: V. Panchal, R. Pearce, R. Yakimova, A. Tzalenchuk and O. Kazakova, ‘Standardization of surface potential measurements of graphene domains’, arXiv:1305.0465 (NPL) (fetch-tool summary); L. Polak and R. J. Wijngaarden, ‘Two competing interpretations of Kelvin probe force microscopy on semiconductors put to test’, arXiv:1605.05539: V_CPD = (W_s - W_p)/e with feedback voltage positive when applied to the sample, surface band bending from surface states or fixed charge, Kelvin-voltage difference across Si p-n junctions versus doping, surface dipoles from termination or adsorbed molecular layers, ions from the environment depositing fixed surface charge, measurements in air (fetch-tool summary).
Ques: Does Bruker’s 240 mV Sn-Pb contrast match their work functions?
Not exactly. Bruker reports 240 with PeakForce KPFM and 235 with FM-KPFM on a Sn-Pb alloy, describing these as consistent with work functions of 4.42 for Sn and 4.25 for Pb, which differ by only 170 ; the measured contrast is 41 per cent larger (computed). The tabulated values, Sn 4.42 and Pb 4.25 (LibreTexts), are for clean metals; work functions vary with crystal facet and contamination can shift measured values by up to 0.5 (Tran et al.), and both metals oxidize in air (my reading). The agreement between PeakForce and FM, 240 and 235 , suggests the measurement is reliable and the reference values are the less likely to apply to this surface (my reading). Absolute work functions need clean, calibrated surfaces. Reference: Bruker, ‘PeakForce Kelvin Probe Force Microscopy’, application note AN140 (several equations garbled) (fetch-tool summary); LibreTexts Chemistry, ‘B1: Workfunction Values (Reference Table)’, no original source named: Au 5.1 to 5.47, Al 4.06 to 4.26, Sn 4.42, Pb 4.25, Os 5.93, Cs 2.1 eV (fetch-tool summary); R. Tran, X.-G. Li, J. H. Montoya, D. Winston, K. A. Persson and S. P. Ong, ‘Anisotropic work function of elemental crystals’, Surf. Sci. 687, 48 (2019): work function varies with crystal facet, surface contamination can cause errors up to 0.5 eV in measured values (fetch-tool summary).
Ques: How repeatable is KPFM between probes?
To about 20 on a good reference. Bruker measured an Au/Al standard, 50 films on n-type silicon, with nine PFQNE-AL probes: Au at 0.639 plus or minus 0.018 , Al at -0.185 plus or minus 0.020 , and the Au-Al difference at 0.825 plus or minus 0.019 . Tabulated work functions, Au 5.1 to 5.47 and Al 4.06 to 4.26 (LibreTexts), give at least 0.84 , so the measurement sits 2 per cent below the smallest tabulated difference (computed). Probe-to-probe scatter of 0.018 is 2.8 per cent of the Au value (computed). Bruker’s text gives deviations below 50 in one place and below 20 in another; the table supports the 20 figure. Differences measured with one tip are more repeatable than absolute values. Reference: Bruker, ‘PeakForce Kelvin Probe Force Microscopy’, application note AN140 (several equations garbled) (fetch-tool summary); LibreTexts Chemistry, ‘B1: Workfunction Values (Reference Table)’, no original source named: Au 5.1 to 5.47, Al 4.06 to 4.26, Sn 4.42, Pb 4.25, Os 5.93, Cs 2.1 eV (fetch-tool summary).
Ques: What else can KPFM measure?
Photovoltages, high surface potentials and nanoscale charge. Bruker reports a work-function down-shift of 535 on an organic solar cell under 300 suns, a single carbon nanotube 30 above its substrate and a bundle 105 , high-voltage KPFM to plus or minus 200 with under 15 per cent error and up to 1000 , and charge on PDMS from -99 to 156 . Bias frequencies must stay below half the cantilever resonance in the high-voltage mode. A 30 nanotube signal is only 0.8 per cent of the maximum possible CPD (computed), showing the sensitivity required. Bruker also uses a glove box with about 1 ppm oxygen and water to avoid adsorbates, which Panchal et al. show alter measured potentials. Reference: Bruker, ‘PeakForce Kelvin Probe Force Microscopy’, application note AN140 (several equations garbled) (fetch-tool summary); V. Panchal, R. Pearce, R. Yakimova, A. Tzalenchuk and O. Kazakova, ‘Standardization of surface potential measurements of graphene domains’, arXiv:1305.0465 (NPL) (fetch-tool summary).
References
- Bruker, “PeakForce Kelvin Probe Force Microscopy,” application note AN140, hosted by Boise State University. Available: https://www.boisestate.edu/wp-content/uploads/sites/551/2019/07/Bruker-AN-140-PeakForce-KPFM.pdf
- L. Polak and R. J. Wijngaarden, “Two competing interpretations of Kelvin probe force microscopy on semiconductors put to test,” arXiv:1605.05539, 2016. Available: https://arxiv.org/pdf/1605.05539
- LibreTexts Chemistry, “B1: Workfunction Values (Reference Table).” Available: https://chem.libretexts.org/Ancillary_Materials/Reference/Reference_Tables/Bulk_Properties/B1:_Workfunction_Values_(Reference_Table)
- R. Tran, X.-G. Li, J. H. Montoya, D. Winston, K. A. Persson and S. P. Ong, “Anisotropic work function of elemental crystals,” Surface Science, vol. 687, pp. 48-55, 2019. Available: https://perssongroup.lbl.gov/papers/tran-2019-work-function.pdf
- V. Panchal, R. Pearce, R. Yakimova, A. Tzalenchuk and O. Kazakova, “Standardization of surface potential measurements of graphene domains,” arXiv:1305.0465. Available: https://arxiv.org/pdf/1305.0465