1. Theory
- Definition: A scanning microwave impedance microscope sends a low-power microwave signal, typically near 3 , to a shielded AFM tip and measures the reflected signal, which depends on the local tip-sample admittance (Bruker; EDFA).
- Two channels: The imaginary part (sMIM-C) follows capacitance and permittivity; the real part (sMIM-R) follows loss and conductivity (Bruker; EDFA).
- Near field: The tip is far smaller than the wavelength, so the interaction is quasi-static and confined to a volume set by the tip size (Zhong et al.; Lai et al.).
- Carrier profiling: Adding an AC sample bias gives dC/dV and dR/dV, like SCM but without needing sample grounding (Bruker; EDFA).
2. Working Principle
- Probe: A coaxially shielded cantilever tip of radius below 50 reduces stray capacitance (Bruker).
- Matching: An impedance-matching network or quarter-wave resonator couples the tip to a 50 line (Zhong et al.; EDFA).
- Demodulation: The reflected wave is mixed down into in-phase and quadrature outputs giving sMIM-R and sMIM-C (Bruker; Lai et al.).
- Scanning: Contact or PeakForce tapping at forces below 100 maps topography and both channels together (Bruker).
3. Applications
- Dopant mapping: SRAM, IGBT and image-sensor cross-sections over to (Bruker; Zhong et al.).
- Failure analysis: Locating gate-oxide point defects or contamination and dopant defects invisible to electron and optical microscopy (EDFA).
- Buried and dielectric structures: Subsurface oxide under nitride, permittivity quantification (Bruker; Zhong et al.).
- Nanomaterials: Carbon nanotubes, nanoparticles and phase-change nanoribbons (Bruker; Zhong et al.).
Admittance and Response
| Name | Formula | Related Formulae | Variables/Constants |
|---|---|---|---|
| Tip-sample admittance | |||
| Lumped response to sample resistance | |||
| Quasi-static and penetration conditions |
Signal Chain and Sensitivity
| Name | Formula | Related Formulae | Variables/Constants |
|---|---|---|---|
| Microwave delivery and power | |||
| Capacitance sensitivity | |||
| Carrier profiling channels |
Resolution and Imaging Conditions
| Name | Formula | Related Formulae | Variables/Constants |
|---|---|---|---|
| Lateral resolution | |||
| PeakForce sMIM force | |||
| Subsurface sensing |
Important Questions
Ques: What does sMIM measure?
The local admittance between tip and sample at microwave frequency. Bruker describes a low-power 3 signal sent to the tip, whose reflection carries the electrodynamic properties of the material beneath it, read as sMIM-C and sMIM-R channels related to permittivity and conductivity. The EDFA article describes sMIM-C as relative capacitance and sMIM-R as relative resistivity of the volume under the tip. Writing the tip admittance, which Lai et al. compute as , as (my reading), a 0.075 capacitance change at 3 is an admittance of only 1.4 (computed), which is why matched, shielded probes are used. The tip is two million times smaller than the 10 wavelength (computed), so the interaction is quasi-static, as Zhong et al. note. Reference: Z. Huang, B. Drevniok, P. De Wolf, S. Dixon-Warren, O. Amster, S. Friedman, B. Pittenger, C. Li and Y. Yang, ‘Nanoscale Mapping of Permittivity and Conductivity with Scanning Microwave Impedance Microscopy’, Bruker Application Note AN145 (fetch-tool summary); ‘Scanning microwave impedance microscopy’ article excerpt, Electronic Device Failure Analysis 25(1), February 2023, page 14 (fetch-tool summary; title and authors not shown); K. Lai, W. Kundhikanjana, M. Kelly and Z.-X. Shen, ‘Modeling of a cantilever-based near-field scanning microwave microscope’, arXiv:0801.2137 (Rev. Sci. Instrum. 79, 063703, 2008): admittance Y computed as 1/Z by quasi-static finite-element analysis at 1 GHz, lumped-element probe description valid since the chip is far smaller than the 30 cm wavelength, imaginary contrast scaling roughly with tip diameter, lambda/4 resonator critically coupled to 50 ohm by a tuning stub, IQ mixer with outputs 90 degrees apart giving capacitive and resistive channels, sensitivity down to the aF level, 110 dB gain at 1 kHz bandwidth, 15 mV per aF (fetch-tool summary); Z. Zhong, X. Chen, X. Quan, H. Huan, F. Nian, S. Liang and Y. Yang, ‘Developments and Recent Progresses in Microwave Impedance Microscope’, Front. Phys. 8, 593076 (2020) (fetch-tool summary).
Ques: Why does the sMIM-R signal peak at intermediate conductivity?
Because loss is largest when the sample resistance matches the coupling reactance. In a lumped model the tip couples through capacitance to a sample resistance , giving (my reading; Lai et al. treat the probe with lumped elements). Insulators ( large) give little loss; metals ( small) short the capacitor and give large but lossless capacitance. The real part peaks at , while sMIM-C rises monotonically with conductivity. For an assumed 1 and 3 , the peak falls at 53 , about 0.53 for a 25 contact (computed estimate). The two channels together therefore resolve the ambiguity of a single peaked curve (my reading). Reference: K. Lai, W. Kundhikanjana, M. Kelly and Z.-X. Shen, ‘Modeling of a cantilever-based near-field scanning microwave microscope’, arXiv:0801.2137 (Rev. Sci. Instrum. 79, 063703, 2008): admittance Y computed as 1/Z by quasi-static finite-element analysis at 1 GHz, lumped-element probe description valid since the chip is far smaller than the 30 cm wavelength, imaginary contrast scaling roughly with tip diameter, lambda/4 resonator critically coupled to 50 ohm by a tuning stub, IQ mixer with outputs 90 degrees apart giving capacitive and resistive channels, sensitivity down to the aF level, 110 dB gain at 1 kHz bandwidth, 15 mV per aF (fetch-tool summary).
Ques: How sensitive is sMIM to capacitance: 0.075 aF or 0.7 fF?
The values come from different probes and eras. The EDFA article reports a measured capacitance sensitivity of 0.075 , and Bruker states sub-aF sensitivity with coaxially shielded probes, about ten times better than traditional techniques. Zhong et al. review earlier MIM probes with minimum sensitivities of 0.7 for Ti/Al-coated nanowire and Pt probes, 3 for a bare nanowire and 6 for a silicon probe, about 9300 times worse than 0.075 (computed). For current shielded commercial probes, the sub-aF figures are more likely, while the femtofarad values describe unshielded or prototype tips where stray capacitance dominates (my reading). Lai et al. quote aF-level sensitivity, 15 per aF, at a 1 bandwidth, so values should be compared at the same bandwidth. Reference: ‘Scanning microwave impedance microscopy’ article excerpt, Electronic Device Failure Analysis 25(1), February 2023, page 14 (fetch-tool summary; title and authors not shown); Z. Huang, B. Drevniok, P. De Wolf, S. Dixon-Warren, O. Amster, S. Friedman, B. Pittenger, C. Li and Y. Yang, ‘Nanoscale Mapping of Permittivity and Conductivity with Scanning Microwave Impedance Microscopy’, Bruker Application Note AN145 (fetch-tool summary); Z. Zhong, X. Chen, X. Quan, H. Huan, F. Nian, S. Liang and Y. Yang, ‘Developments and Recent Progresses in Microwave Impedance Microscope’, Front. Phys. 8, 593076 (2020) (fetch-tool summary); K. Lai, W. Kundhikanjana, M. Kelly and Z.-X. Shen, ‘Modeling of a cantilever-based near-field scanning microwave microscope’, arXiv:0801.2137 (Rev. Sci. Instrum. 79, 063703, 2008): admittance Y computed as 1/Z by quasi-static finite-element analysis at 1 GHz, lumped-element probe description valid since the chip is far smaller than the 30 cm wavelength, imaginary contrast scaling roughly with tip diameter, lambda/4 resonator critically coupled to 50 ohm by a tuning stub, IQ mixer with outputs 90 degrees apart giving capacitive and resistive channels, sensitivity down to the aF level, 110 dB gain at 1 kHz bandwidth, 15 mV per aF (fetch-tool summary).
Ques: What lateral resolution does sMIM achieve?
Sources range from about 30 to about 1 . Bruker states that sMIM easily achieves electrical resolution below 30 , below 20 in its conclusion, and resolved a p-to-n transition narrower than 10 on an SRAM, with tips below 50 radius. The EDFA article reports resolution below the tip diameter because sMIM is a near-field measurement, and about 1 measured on a Moire pattern, 50 times finer than a 50 radius (computed). The 10 to 30 range is more likely for routine semiconductor cross-sections, while 1 applies to periodic contrast on ideal samples, where a sharp asperity carries most of the signal (my reading). Reference: Z. Huang, B. Drevniok, P. De Wolf, S. Dixon-Warren, O. Amster, S. Friedman, B. Pittenger, C. Li and Y. Yang, ‘Nanoscale Mapping of Permittivity and Conductivity with Scanning Microwave Impedance Microscopy’, Bruker Application Note AN145 (fetch-tool summary); ‘Scanning microwave impedance microscopy’ article excerpt, Electronic Device Failure Analysis 25(1), February 2023, page 14 (fetch-tool summary; title and authors not shown).
Ques: How does sMIM profile dopants, and how does it compare with SCM?
By measuring dC/dV with an AC sample bias, as SCM does, plus resistance channels. Bruker notes that with AC sample bias sMIM provides SCM-like carrier profiling and adds a dR/dV channel for nonlinear resistive behaviour, without electrical contact to the substrate. The EDFA article states that the dC/dV phase indicates carrier type and the amplitude relates to doping, and that sMIM needs no sample grounding and collects R and C together. Bruker reports dopant sensitivity from to on image sensors, six decades (computed), broader than the factor of 2 signal change of SCM over four decades on the SCM page. Zhong et al. cite SRAM and staircase imaging. Reference: Z. Huang, B. Drevniok, P. De Wolf, S. Dixon-Warren, O. Amster, S. Friedman, B. Pittenger, C. Li and Y. Yang, ‘Nanoscale Mapping of Permittivity and Conductivity with Scanning Microwave Impedance Microscopy’, Bruker Application Note AN145 (fetch-tool summary); ‘Scanning microwave impedance microscopy’ article excerpt, Electronic Device Failure Analysis 25(1), February 2023, page 14 (fetch-tool summary; title and authors not shown); Z. Zhong, X. Chen, X. Quan, H. Huan, F. Nian, S. Liang and Y. Yang, ‘Developments and Recent Progresses in Microwave Impedance Microscope’, Front. Phys. 8, 593076 (2020) (fetch-tool summary).
Ques: Can sMIM image soft or buried structures?
Yes, with PeakForce tapping and its subsurface sensitivity. Bruker combines sMIM with PeakForce tapping at 1 to 2 , peak forces routinely below 100 or even 50 , and contact times of tens to hundreds of microseconds, imaging carbon nanotubes and nanoparticles. That force is about 33000 times below the 3.3 of SSRM (SSRM page; computed). Because the microwave field penetrates into the stack, Bruker detected SiO2 with permittivity 3.9 buried 133 under Si3N4 of permittivity 7.5, a ratio of 1.9 (computed), on a surface with 0.4 roughness. Zhong et al. cite imaging of buried silicon donor patterns. Zhong et al. note the strong penetration of microwaves, which lets sMIM probe internal structure. Reference: Z. Huang, B. Drevniok, P. De Wolf, S. Dixon-Warren, O. Amster, S. Friedman, B. Pittenger, C. Li and Y. Yang, ‘Nanoscale Mapping of Permittivity and Conductivity with Scanning Microwave Impedance Microscopy’, Bruker Application Note AN145 (fetch-tool summary); Z. Zhong, X. Chen, X. Quan, H. Huan, F. Nian, S. Liang and Y. Yang, ‘Developments and Recent Progresses in Microwave Impedance Microscope’, Front. Phys. 8, 593076 (2020) (fetch-tool summary).
Ques: How is microwave power delivered to the tip?
Through a matched, shielded transmission path at low power. Bruker uses a default -20 , 10 , adjustable from -10 to -40 , 0.1 to 100 (computed), with the optimal frequency a few to tens of from 3 to match the probe. Zhong et al. describe a tip connected to a quarter-wave section of transmission line forming a resonator, critically coupled to a 50 feed line through an open-end tuning stub; in an assumed PTFE line that quarter wave is about 1.7 (computed). The EDFA article describes a custom cantilever with matched impedance and shielding. Zhong et al. note that without matching the large mismatch to the 50 line makes small impedance variations undetectable. Reference: Z. Huang, B. Drevniok, P. De Wolf, S. Dixon-Warren, O. Amster, S. Friedman, B. Pittenger, C. Li and Y. Yang, ‘Nanoscale Mapping of Permittivity and Conductivity with Scanning Microwave Impedance Microscopy’, Bruker Application Note AN145 (fetch-tool summary); Z. Zhong, X. Chen, X. Quan, H. Huan, F. Nian, S. Liang and Y. Yang, ‘Developments and Recent Progresses in Microwave Impedance Microscope’, Front. Phys. 8, 593076 (2020) (fetch-tool summary); ‘Scanning microwave impedance microscopy’ article excerpt, Electronic Device Failure Analysis 25(1), February 2023, page 14 (fetch-tool summary; title and authors not shown); K. Lai, W. Kundhikanjana, M. Kelly and Z.-X. Shen, ‘Modeling of a cantilever-based near-field scanning microwave microscope’, arXiv:0801.2137 (Rev. Sci. Instrum. 79, 063703, 2008): admittance Y computed as 1/Z by quasi-static finite-element analysis at 1 GHz, lumped-element probe description valid since the chip is far smaller than the 30 cm wavelength, imaginary contrast scaling roughly with tip diameter, lambda/4 resonator critically coupled to 50 ohm by a tuning stub, IQ mixer with outputs 90 degrees apart giving capacitive and resistive channels, sensitivity down to the aF level, 110 dB gain at 1 kHz bandwidth, 15 mV per aF (fetch-tool summary).
References
- Z. Huang, B. Drevniok, P. De Wolf, S. Dixon-Warren, O. Amster, S. Friedman, B. Pittenger, C. Li and Y. Yang, “Nanoscale Mapping of Permittivity and Conductivity with Scanning Microwave Impedance Microscopy,” Bruker Application Note AN145. Available: https://bruker.com/es/products-and-solutions/microscopes/materials-afm/resource-library/an145-nanoscale-mapping-of-permittivity-and-conductivity-with-scanning-microwave-impedance-microscopy.html
- “Scanning microwave impedance microscopy” (article excerpt), Electronic Device Failure Analysis, vol. 25, no. 1, p. 14, February 2023. Available: http://static.asminternational.org/EDFA/202302/10/
- K. Lai, W. Kundhikanjana, M. Kelly and Z.-X. Shen, “Modeling and characterization of a cantilever-based near-field scanning microwave impedance microscope,” Review of Scientific Instruments, vol. 79, 063703, 2008; preprint arXiv:0801.2137. Available: https://www.arxiv.org/pdf/0801.2137
- Z. Zhong, X. Chen, X. Quan, H. Huan, F. Nian, S. Liang and Y. Yang, “Developments and Recent Progresses in Microwave Impedance Microscope,” Frontiers in Physics, vol. 8, 593076, 2020. Available: https://frontiersin.org/journals/physics/articles/10.3389/fphy.2020.593076/full
Related pages in this vault: Scanning Spreading Resistance Microscope (SSRM), Atomic Force Microscope (AFM), Scanning Capacitance Microscope (SCM), Scanning Near-Field Optical Microscope (SNOM) and TERS, Kelvin Probe Force Microscope (KPFM).