1. Theory

  • Definition: Critical-dimension small-angle X-ray scattering (CD-SAXS) transmits a hard X-ray beam through a wafer and records diffraction from periodic line gratings, giving pitch, width, height, sidewall angle and roughness averaged over many lines (Hu et al.; Wu et al.; Sunday et al.).
  • Fourier relation: Scattered intensity is the squared Fourier transform of the electron-density profile, damped by a Debye-Waller roughness factor (Sunday et al.).
  • Reciprocal space: A grating of pitch P produces orders spaced in ; their intensity along encodes the cross-section shape (Hu et al.).
  • Rotation: Tilting the sample sweeps the orders through , revealing height and sidewall angle (Hu et al.; Sunday et al.).

2. Working Principle

  • Source: Hard X-rays above about 15 from a synchrotron or high-brightness laboratory source pass through the silicon wafer (Sunday et al.; Wu et al.).
  • Detection: A 2-D area detector several metres downstream records the diffraction orders (Hu et al.).
  • Tilt series: Images at many rotation angles, for example -20 to 20 or over 45 in 1 steps, build the - map (Hu et al.; Sunday et al.).
  • Modelling: Stacked-trapezoid models are fitted, with uncertainties from Markov-chain Monte Carlo (Sunday et al.).

3. Applications

  • Lithography focus and dose control: Width, height and sidewall angle across focus series (Wu et al.).
  • Dense patterns: Sub-50 pitch gratings, fins and self-aligned multiple patterning (Sunday et al.).
  • Reference metrology: Validation of CD-SEM and optical scatterometry with TEM cross sections (Sunday et al.; Hu et al.).
  • Line roughness: Averaged sidewall and top-surface roughness (Wu et al.).

Scattering Geometry

NameFormulaRelated FormulaeVariables/Constants
Scattering vector



Grating pitch from peak spacing

Intensity model


Profile Reconstruction

NameFormulaRelated FormulaeVariables/Constants
Trapezoid form factor


Height and profile from sample rotation

Sidewall angle and width results

Practical Limits

NameFormulaRelated FormulaeVariables/Constants
Transmission through the wafer


Source flux and measurement time

Roughness: SAXS against SEM

Important Questions

Ques: How does CD-SAXS measure the pitch of a grating?

From the spacing of diffraction orders in reciprocal space. The scattering vector is , and a grating of pitch P gives orders at (Hu et al.; Sunday et al.; background knowledge, Als-Nielsen and McMorrow). For Hu’s 0.729 beam, 17.0 (computed), a 32 pitch produces orders 0.196 apart, 2.28 in angle and 15.9 apart on a detector 7 away, about 202 pixels (computed). Wu et al. report pitches of 237.1 +/- 0.5 and 330.5 +/- 0.5 , 0.15 per cent precision (computed), and sub-nanometre precision generally. Because pitch comes from peak positions alone, it is independent of the shape model, unlike width and sidewall angle. Reference: T. Hu, R. L. Jones, W.-L. Wu, E. K. Lin, Q. Lin and J. Quintana, ‘Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Gratings’, IBM Research Report RC23206 (2004) (fetch-tool summary); D. F. Sunday, D. Windover, W. Wu and R. J. Kline, ‘Current Status of CDSAXS: Is it Fab-Ready?’, NIST MSED slides (fetch-tool summary; partly garbled); J. Als-Nielsen and D. McMorrow, Elements of Modern X-ray Physics, 2nd ed., Wiley 2011, and NIST XCOM attenuation data (background knowledge; not opened); W.-L. Wu, R. L. Jones, T. Hu, C. L. Soles and E. K. Lin, ‘Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Line Gratings’, ULSI conference slides, Richardson TX (2005) (fetch-tool summary).

Ques: How do sample rotations reveal height and sidewall angle?

By sweeping each order through . At rotation , an order at samples , so a tilt series maps the form factor along (background knowledge, Als-Nielsen and McMorrow). Fringes spaced give height: Hu et al. found about 318 , a period of 0.0198 (computed). For a 400 line at 32 pitch, one fringe needs about 4.6 of rotation (computed), so Sunday’s 45 range in 1 steps samples many fringes. A sloped sidewall tilts the intensity streaks, and Hu et al. obtained sidewall angles from linear fits of the major maxima, rotating from -20 to 20 . Sunday et al. fit six stacked trapezoids with Monte Carlo uncertainty estimates. Reference: T. Hu, R. L. Jones, W.-L. Wu, E. K. Lin, Q. Lin and J. Quintana, ‘Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Gratings’, IBM Research Report RC23206 (2004) (fetch-tool summary); D. F. Sunday, D. Windover, W. Wu and R. J. Kline, ‘Current Status of CDSAXS: Is it Fab-Ready?’, NIST MSED slides (fetch-tool summary; partly garbled); J. Als-Nielsen and D. McMorrow, Elements of Modern X-ray Physics, 2nd ed., Wiley 2011, and NIST XCOM attenuation data (background knowledge; not opened).

Ques: Is the sidewall angle of Hu’s sample A 5.7 or 5.0 degrees?

Hu et al. report 5.7 +/- 0.2 from linear fits of the major maxima and about 5.0 from fitting the second-order peak intensity. The maxima-tilt method depends only on the direction of the streaks and is less sensitive to roughness and intensity calibration, so 5.7 is more likely, with the intensity fit pulled lower by roughness damping that the simple model omits (my reading). Wu’s later deck shows similar scatter: angles of 2 +/- 0.5 and 5.6 +/- 0.5 across a focus series, with uncertainties stated inconsistently. A 3 Debye-Waller roughness already reduces order 2 at 32 pitch to 25 per cent (computed), so intensity-based angles need roughness in the model. Hu et al. found qualitative agreement with tilted SEM cross sections. Reference: T. Hu, R. L. Jones, W.-L. Wu, E. K. Lin, Q. Lin and J. Quintana, ‘Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Gratings’, IBM Research Report RC23206 (2004) (fetch-tool summary); W.-L. Wu, R. L. Jones, T. Hu, C. L. Soles and E. K. Lin, ‘Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Line Gratings’, ULSI conference slides, Richardson TX (2005) (fetch-tool summary).

Ques: Why do synchrotron and laboratory CD-SAXS times differ so much?

Because of incident flux. Wu et al. used a synchrotron spot of about 40 by 40 with 1 to 5 s per sample, whereas Sunday et al. report a laboratory source of about in a 300 beam, needing 121 images of 10 s, about 30 minutes, per scan. Their ideal fab source needs to for a 10 s measurement, 45 to 4500 times more flux (computed), and a spot below 100 to fit test pads. The wafer itself also absorbs: a 775 silicon wafer transmits 15, 45 and 77 per cent at 15, 20 and 30 (computed from background attenuation data, NIST XCOM), and Hu’s 1 at 17 , 18 per cent, matches their 20 per cent (computed). Reference: W.-L. Wu, R. L. Jones, T. Hu, C. L. Soles and E. K. Lin, ‘Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Line Gratings’, ULSI conference slides, Richardson TX (2005) (fetch-tool summary); D. F. Sunday, D. Windover, W. Wu and R. J. Kline, ‘Current Status of CDSAXS: Is it Fab-Ready?’, NIST MSED slides (fetch-tool summary; partly garbled); J. Als-Nielsen and D. McMorrow, Elements of Modern X-ray Physics, 2nd ed., Wiley 2011, and NIST XCOM attenuation data (background knowledge; not opened); T. Hu, R. L. Jones, W.-L. Wu, E. K. Lin, Q. Lin and J. Quintana, ‘Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Gratings’, IBM Research Report RC23206 (2004) (fetch-tool summary).

Ques: How does CD-SAXS roughness compare with SEM line-edge roughness?

They measure different things. Wu et al. state that CD-SAXS roughness includes both sidewall and top surface, unlike SEM line-edge roughness, which follows only the edge; they report 3 to 5 RMS for a high-LER resist. In SAXS, roughness enters through the Debye-Waller factor (Sunday et al.), suppressing higher orders: at 32 pitch, 1 roughness leaves 38 per cent of order 5 (computed), so rough lines show few orders. SAXS averages over every line in the beam, about 1250 lines and 5 cm of total edge length for a 40 spot at 32 pitch (computed), while SEM samples short lengths of a few lines (my reading). Hu et al. note LER as one cause of faster-than-expected intensity decay. Reference: W.-L. Wu, R. L. Jones, T. Hu, C. L. Soles and E. K. Lin, ‘Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Line Gratings’, ULSI conference slides, Richardson TX (2005) (fetch-tool summary); D. F. Sunday, D. Windover, W. Wu and R. J. Kline, ‘Current Status of CDSAXS: Is it Fab-Ready?’, NIST MSED slides (fetch-tool summary; partly garbled); T. Hu, R. L. Jones, W.-L. Wu, E. K. Lin, Q. Lin and J. Quintana, ‘Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Gratings’, IBM Research Report RC23206 (2004) (fetch-tool summary).

References

  1. T. Hu, R. L. Jones, W.-L. Wu, E. K. Lin, Q. Lin and J. Quintana, “Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Gratings,” IBM Research Report RC23206, 2004. Available: https://dominoweb.draco.res.ibm.com/reports/rc23206.pdf
  2. D. F. Sunday, D. Windover, W. Wu and R. J. Kline, “Current Status of CDSAXS: Is it Fab-Ready?,” NIST Materials Science and Engineering Division slides. Available: https://www.nist.gov/document/klinepdf
  3. J. Als-Nielsen and D. McMorrow, Elements of Modern X-ray Physics, 2nd ed., Wiley, 2011; NIST XCOM photon cross-section database (cited from background knowledge).
  4. W.-L. Wu, R. L. Jones, T. Hu, C. L. Soles and E. K. Lin, “Small Angle X-ray Scattering Metrology for Sidewall Angle and Cross Section of Nanometer Scale Line Gratings,” ULSI conference, Richardson, TX, 2005. Available: https://www.nist.gov/document/wu2005pdf

Related pages in this vault: X-Ray Reflectometer (XRR), CD-SEM, X-Ray Diffractometer (XRD and HRXRD), Optical CD Scatterometer (OCD), Spectroscopic Ellipsometer, Transmission Electron Microscope (TEM and STEM), X-Ray Topography, Scanning Electron Microscope (SEM), Rutherford Backscattering Spectrometer (RBS).