1. Theory
- Definition: E-beam inspection scans a patterned wafer with a focused electron beam to find physical and electrical defects, and a defect-review SEM revisits defects found by inspection to image and classify them at high magnification (low-voltage patent; Hitachi patent).
- Voltage contrast: At landing energies between the two crossover energies, floating conductors charge differently from grounded ones, so opens and shorts below the surface change the secondary-electron signal (low-voltage patent; Applied Materials patent).
- Physical contrast: Topography and material contrast reveal residues, bridges and missing patterns at nanometre resolution (Scanning Electron Microscope (SEM) page).
- Speed limit: Shot noise sets the electrons needed per pixel, so beam current limits area per hour (Du and Jacobsen; KLA patent).
2. Working Principle
- Column: Field-emission source, low landing energy set by wafer bias, energy-filtered secondary detection (low-voltage patent).
- Charge control: Flood or defocused pre-charge before imaging sets the surface potential (low-voltage patent; Applied Materials patent).
- Multi-beam: Arrays of beamlets with separate detectors raise total current (KLA patent).
- Review: Low-magnification relocation, defect detection against a reference, high-magnification imaging, then classification (Hitachi patent).
3. Applications
- Contacts and vias: Under-etched or unlanded vias and contact opens found by voltage contrast (Applied Materials patent; low-voltage patent).
- Gate and metal shorts: Leakage paths and bridges in logic and memory arrays (low-voltage patent).
- Defect review and Pareto: Imaging and classifying optical inspection hits (Hitachi patent).
Charging and Voltage Contrast
| Name | Formula | Related Formulae | Variables/Constants |
|---|---|---|---|
| Total electron yield and charging | |||
| Landing energy with wafer bias | |||
| Floating-conductor charging |
Signal, Noise and Throughput
| Name | Formula | Related Formulae | Variables/Constants |
|---|---|---|---|
| Rose criterion and electrons per pixel | |||
| Pixel rate | |||
| Area rate and wafer time |
Pre-Charge and Defect Review
| Name | Formula | Related Formulae | Variables/Constants |
|---|---|---|---|
| Pre-charge dose | |||
| Defect relocation window | |||
| Reference image and classification |
Important Questions
Ques: How does voltage contrast reveal a buried electrical defect?
By making connected and floating structures charge to different potentials. The low-voltage patent operates between the two crossover energies, where each primary electron releases more than one secondary, so an isolated conductor charges positive while a grounded one stays near zero. Thomson et al. note that a few volts of positive potential pull most secondaries back to the surface, so a positive conductor appears darker (my reading), so an unlanded via or open contact stands out from its connected neighbours. The Applied Materials patent states the requirement as a charging difference between the defective structure and its surroundings, larger than the secondary-electron energy width if an energy filter separates them. A 0.1 floating plug needs only 624 electrons to shift by 1 , about 10 at 10 net current (assumed; computed). Reference: ‘Voltage contrast method and apparatus for semiconductor inspection using low voltage particle beam’, US patent 6,566,897 (divisional of US 6,344,750, filed January 1999) (fetch-tool summary); E. T. Bullock, ‘Apparatus and method for enhancing voltage contrast of a wafer’, US patent 7,994,476, Applied Materials Israel (filed November 2007) (fetch-tool summary); C. D. Thomson, V. Zavyalov, J. R. Dennison and J. Corbridge, ‘Electron Emission Properties of Insulator Materials Pertinent to the International Space Station’, Utah State University (NASA NTRS), 2003: SE and BSE yields, unity total yield at the first (below 100 V) and second (above 1 keV) crossover energies, net positive charging between them, a few-volt positive potential pulling most SEs back to the surface (fetch-tool summary).
Ques: Why are landing energies near 1 keV used for voltage contrast?
Because charging polarity depends on the total yield . Between the crossover energies and the yield exceeds one, and the surface charges positively (Thomson et al.); an insulator then reaches a steady state where the total yield equals one. The low-voltage patent gives landing energies of 500 to 1.5 in this window, set partly by wafer bias, and notes that thin silicide needs low landing energy to avoid beam punch-through leaking the charge. With an assumed 10 column and a wafer bias of -9.5 , the landing energy is 0.5 (computed). Above the surface charges negatively instead, which is the negative voltage-contrast mode the patent also describes. Reference: C. D. Thomson, V. Zavyalov, J. R. Dennison and J. Corbridge, ‘Electron Emission Properties of Insulator Materials Pertinent to the International Space Station’, Utah State University (NASA NTRS), 2003: SE and BSE yields, unity total yield at the first (below 100 V) and second (above 1 keV) crossover energies, net positive charging between them, a few-volt positive potential pulling most SEs back to the surface (fetch-tool summary); ‘Voltage contrast method and apparatus for semiconductor inspection using low voltage particle beam’, US patent 6,566,897 (divisional of US 6,344,750, filed January 1999) (fetch-tool summary).
Ques: Why does e-beam inspection sample instead of scanning whole wafers?
Because shot noise sets a minimum number of electrons per pixel. The Rose criterion with contrast 0.1, SNR 5 and half of the primaries detected gives 5000 primaries per pixel (assumed; computed). At 100 this takes 8.0 , a 125 pixel rate, just above the 1 to 100 range of the low-voltage patent (computed). With 10 pixels one beam covers 45 , so a full 300 wafer would take 1573 , about 66 days (computed). Inspection is therefore restricted to sampled dies, arrays or hot-spot locations, and larger pixels are used for voltage contrast, where the signal is strong (my reading). Higher current enlarges the spot through aberrations and electron interactions. Reference: M. Du and C. Jacobsen, ‘Relative merits and limiting factors for x-ray and electron microscopy of thick, hydrated organic materials’, arXiv:2004.10069 (Ultramicroscopy 184, 2018): Rose criterion SNR = 5, n = SNR^2/Theta^2 quanta per pixel (fetch-tool summary); ‘Voltage contrast method and apparatus for semiconductor inspection using low voltage particle beam’, US patent 6,566,897 (divisional of US 6,344,750, filed January 1999) (fetch-tool summary).
Ques: How do multi-beam systems change e-beam inspection throughput?
By dividing current among many beamlets, each with its own spot and detector. The KLA patent uses 25 beamlets in a 5 by 5 array, with a stated range of 2 to 200, and claims more than 10 and up to 25 times the total current of a single beam at the same spot size, since lens aberrations, source brightness and electron interactions limit one beam. It cites a roadmap need for more than 0.1 wafer per hour at below three times the design rule. At the assumed 10 pixel, 25 beams reduce a full wafer from 1573 to 63 , and about 157 beams would be needed for 10 (computed), so full-wafer coverage at small pixels still requires sampling (my reading). Reference: ‘Multiple-beam system for high-speed electron-beam inspection’, US patent 8,362,425, KLA-Tencor (filed April 2011) (fetch-tool summary); M. Du and C. Jacobsen, ‘Relative merits and limiting factors for x-ray and electron microscopy of thick, hydrated organic materials’, arXiv:2004.10069 (Ultramicroscopy 184, 2018): Rose criterion SNR = 5, n = SNR^2/Theta^2 quanta per pixel (fetch-tool summary).
Ques: Why is the surface pre-charged before voltage-contrast imaging?
To set a stable potential difference before the imaging scan reads it. The low-voltage patent uses a flood beam with at least four times the imaging current over an area at least twice the image area; its best tabulated setting is 250 by 250 at C in 60 , which is and 3.3 (computed). It also models a micro-retarding field of -0.71 from returning secondaries charging the surroundings. The Applied Materials patent instead defocuses the main beam to at least about 1 , using scan fly-back for pre-charging, controlled by defocus, duty cycle and scan orientation. Both aim for charging differences large enough to survive leakage during imaging (my reading). Reference: ‘Voltage contrast method and apparatus for semiconductor inspection using low voltage particle beam’, US patent 6,566,897 (divisional of US 6,344,750, filed January 1999) (fetch-tool summary); E. T. Bullock, ‘Apparatus and method for enhancing voltage contrast of a wafer’, US patent 7,994,476, Applied Materials Israel (filed November 2007) (fetch-tool summary).
Ques: How does a review SEM find a defect reported by optical inspection?
By imaging a field larger than the coordinate error, then zooming in. The Hitachi patent states that inspection coordinates carry about plus or minus 4 of error, so the review SEM first images at about 15,000 times with a 9 field, which covers the 8 uncertainty span (computed). The defect is located by comparing with a reference image and then reimaged at high magnification for defects of dozens of nanometres. Prior practice took the reference from a die one chip away; the patent synthesizes it from the periodic pattern in the same low-magnification image, avoiding a stage move. With 512 pixels per line, the 9 field gives 17.6 pixels (assumed; computed). Classification of the imaged defect follows. Reference: M. Harada, R. Nakagaki and K. Obara, ‘Method and its apparatus for reviewing defects’, US patent 8,121,397, Hitachi High-Technologies (filed April 2009) (fetch-tool summary).
Ques: What high magnification is used for review: 50,000 or 100,000 times?
The Hitachi patent gives both. Its background describes high-magnification review at about 50,000 times with a roughly 2.5 field, while its embodiment step uses about 100,000 times. Scaling the stated 9 field at 15,000 times gives 2.7 at 50,000 times, consistent with the 2.5 quoted, and 1.35 at 100,000 times (computed). With 512 pixels these give 5.3 and 2.6 pixels (assumed; computed). The patent targets defects of dozens of nanometres needing several-nanometre resolution, so 100,000 times is more likely for the advanced-node case it describes, while 50,000 times reflects general prior practice (my reading). Review recipes store magnification with acceleration voltage, probe current and frame count, so either may appear. Reference: M. Harada, R. Nakagaki and K. Obara, ‘Method and its apparatus for reviewing defects’, US patent 8,121,397, Hitachi High-Technologies (filed April 2009) (fetch-tool summary).
References
- “Voltage contrast method and apparatus for semiconductor inspection using low voltage particle beam,” US patent 6,566,897 (divisional of US application 09/227,395, filed January 8, 1999, now US 6,344,750). Available: https://image-ppubs.uspto.gov/dirsearch-public/print/downloadPdf/6566897
- E. T. Bullock, “Apparatus and method for enhancing voltage contrast of a wafer,” US patent 7,994,476 B2, Applied Materials Israel, filed November 5, 2007. Available: https://patents.google.com/patent/US7994476
- C. D. Thomson, V. Zavyalov, J. R. Dennison and J. Corbridge, “Electron emission properties of insulator materials pertinent to the International Space Station,” Utah State University, 2003 (NASA NTRS 20040111072). Available: https://ntrs.nasa.gov/api/citations/20040111072/downloads/20040111072.pdf
- M. Du and C. Jacobsen, “Relative merits and limiting factors for x-ray and electron microscopy of thick, hydrated organic materials,” Ultramicroscopy, vol. 184, pp. 293-309, 2018; arXiv:2004.10069. Available: https://arxiv.org/abs/2004.10069
- “Multiple-beam system for high-speed electron-beam inspection,” US patent 8,362,425 B2, KLA-Tencor, filed April 27, 2011. Available: https://patents.google.com/patent/US8362425B2/en
- M. Harada, R. Nakagaki and K. Obara, “Method and its apparatus for reviewing defects,” US patent 8,121,397 B2, Hitachi High-Technologies, filed April 1, 2009. Available: https://patents.google.com/patent/US8121397
Related pages in this vault: Patterned-Wafer Optical Inspection, Scanning Electron Microscope (SEM), CD-SEM, Laser-Scattering Particle Inspection, EBIC and EBAC, Focused Ion Beam (FIB), Contacts and Vias, Overlay Metrology Tool, Emission Microscope, OBIRCH and Lock-In Thermography, High-NA DUV Immersion Scanner, Electron-Matter Interaction, Electron Optics and the SEM Column, Defect Detection Fundamentals, Defect Sources and Classification, EUV Scanner, Defect Inspection Workflow and Sampling Plans, Automatic Defect Classification and Machine Learning.